Operating Instructions
MAVOMAX 60, RK1, RK2/RK5
Interior Light Monitor 4/08.17
The MAVOMAX room light monitor is used wherever a constant illuminance has to be ensured. This applies, in
particular, to compliance with the room classes according to DIN 6868-157 of diagnostic rooms, examination
rooms with immediate diagnostics, dental diagnostic workstations and in teleradiology, where diagnostic
monitors and light boxes for medical applications are used. For universally usable rooms with dimmable lighting,
it supports the setting of the permissible illuminance for diagnostic.
During the transitional period for the quality assurance of image reproduction devices according to DIN V 6868-
57, the MAVOMAX can be used in order to extend the interval of constancy testing of image display devices for
veiling luminance and maximum contrast to 6 months. In the case of repeat measurements within the framework
of acceptance or constancy testing, veiling luminance need not be measured again and the 60-minute waiting
time until the image display device has reached a stable state is eliminated.